For the System Safety, please follow the procedure to prolong the lifetime of the tube and reduce any damage to the System
· Set the Main Power Switch of the System to ON position | |
·
Switch on the
power of the Enclosure Safety Interlock
Circuit Box, located in the lower right corner of
the Enclosure base |
NOTICE:
The Green LEDs in the upper right corner of the Enclosure should be lit up, or the Red LEDs
will lit up if one of the doors may not be closed well
·
Switch on the power of Diffractometer
Control Unit and switch on the Heat Exchanger. Press
the PUSH TO START button to circulate the cooling water, set
the temperature knob to 20 | |
·
Press the POWER ON
button on the front panel of the High Tension Generator
| |
·
Wait for 10 minutes
to cool down the re-circulating
water and prepare your sample during this period |
· Set the kV & mA knobs to its minimum reading i.e., 10kV & 10mA | |
· Should the reading of the Re-circulating Temperature reach 20oC, then press the HT ON button to power up the High Tension | |
· Gradually adjust the Voltage Knob to its Max. Value 40kV in a step of 5kV and wait 5seconds every step, then adjust the Current Knob to its Max. Value 30mA, in a step of 5mA and wait 5 sec. per step |
NOTICE:
During Operation, keep an
eye on the reading of Re-circulating Temperature.
Should you find its higher than 22OC, immediately STOP
the scan
and SHUT DOWN the System following
the steps in (IV) Shutting Down the System
· Create a new folder under the directory C:/SIE122/DATA/ for saving your data | |
· Double click the icon Sie1770l.exe on the desktop to run the Scanning Software. From the File Menu, select System and click Software Operation, then from the Dialog Box choose your folder name and click OK | |
· Set the Scanning Parameters for your sample. The maximum 2 theta angle is 100o | |
· Insert the Sample Holder into the Mounting Clamp of the Goniometer, well replace all the Enclosure doors and open the Shutter to start scan. All the data done will be automatically saved under the folder you chose above | |
·
To analyze your
data, run the software Traces
5.2 |
NOTICE:
Before Opening the software Sie1770l.exe, please DON'T
Open the Shutter, or it will damaged the
detector
· After finish scanning and analyzing, close the Shutter and take out the sample, replace all the accessories, clean the desk and Log-off the software and switch off the computer. | |
·
Gradually reduce the Current
in a step of 5mA to its minimum value, then reduce the Voltage in a
step of 5kV to its minimum value, press HT OFF button to
switch off the High Tension Generator | |
·
Let the Heat
Exchanger running 5-10 minutes more to
cool down the X-Ray Tube. | |
·
Now you could switch off all the power of the
System in an alternative order of the steps in
(I) Warming Up the System |